聯系熱線
美國2B臭氧檢測儀106-L與205性能對比
下表比較了各種 2B Tech 臭氧監測儀的規格,旨在幫助確定哪種臭氧監測儀更適合您的應用。如果您有任何疑問或需要幫助來選擇更適合您需求的臭氧監測儀,請隨時與我們聯系。
Specification | Model | ||
---|---|---|---|
202 | 205 | 106-L | |
Measurement Principle | UV Absorbance | UV Absorbance | UV Absorbance |
Dual Beam? | Optional Upgrade | Yes | No |
Federal Equivalent Method (FEM) | Yes | Yes | Yes |
Linear Dynamic Range | 0-250 ppm | 0-250 ppm | 0-100 ppm |
Resolution | 0.1 ppb | 0.1 ppb | 0.1 ppb |
Precision (1σ for 10-s average; aka rms noise) | 1.5 ppb or 2% of Reading | 1.0 ppb or 2% of Reading | 1.5 ppb or 2% of Reading |
Accuracy | 1.5 ppb or 2% of Reading | 1.0 ppb or 2% of Reading | 1.5 ppb or 2% of Reading |
Calibration | NIST Traceable | NIST Traceable | NIST Traceable |
Limit of Detection (10-s measurement, 2σ) | 3.0 ppb | 2.0 ppb | 3.0 ppb |
Flow Rate (nominal) | 1 L/min | 1.8 L/min | 1 L/min |
Flow Rate Requirement | >0.6 L/min | >1.2 L/min | >0.6 L/min |
Baseline Drift | <2 ppb/day <5 ppb/year | <1 ppb/day <3 ppb/year | <3 ppb/day <6 ppb/year |
Sensitivity Drift | <1%/day <3%/year | <1%/day <3%/year | <1%/day <3%/year |
Measurement Time, Frequency | 10 s, 0.1 Hz | 2 s, 0.5 Hz | 2 s, 0.5 Hz |
Response Time, 100% of Step Change | 20 s 2 points | 4 s 2 points | 4 s 2 points |
Averaging Times | 1 min, 5 min, 1 hr | 10 s, 1 min, 5 min, 1 hr | 10 s, 1 min, 5 min, 1 hr |
Ozone Units | ppb, pphm, ppm, μg m-3, mg m-3 | ppb, pphm, ppm, μg m-3, mg m-3 | ppb, pphm, ppm, μg m-3, mg m-3 |
Pressure Units | mbar, torr | mbar, torr | mbar, torr |
Temperature Units | °C, K | °C, K | °C, K |
T and P Corrected | Yes | Yes | Yes |
Operating Temperature Range | 0 to 50 °C; -20 to 50 °C with low T Mods | 0 to 50 °C; -20 to 50 °C with low T Mods | 0 to 50 °C |
Operating Altitude | ~0-13.5 km (0-25 km option) | ~0-13.5 km (0-25 km option) | ~0-13.5 km (0-25 km option) |
Power Requirement | 100-240 VAC, 50/60 Hz; 11-14 VDC, nominally 600 mA at 12 V; 7.2 watt | 100-240 VAC, 50/60 Hz; 11-14 VDC, nominally 650 mA at 12 V; 8 watt | 100-240 VAC, 50/60 Hz; 11-28 VDC, nominally 500 mA at 12 V; 6.0 watt |
Size | 3.7×8.5×11.8 in 9.5×21.5×30 cm | 3.7×8.5×11.5 in 9.5×21.5×30 cm | Standard Case: 3.6×7.9×9.4 in 9x20x24 cm OEM Version: 2.5x7x9 in 6.2×17.8×22.9 cm |
Weight | 5.5 lb (2.5 kg) Without instrument case: 2.8 lb (1.3 kg) | 5.8 lb (2.6 kg) Without instrument case: 3.1 lb (1.4 kg) | Standard Case: 4.2 lb (1.9 kg) OEM Version: 2.8 lb (1.3 kg) |
Data Logger Capacity | 16,383 lines | 16,383 lines | 32,736 lines |
Analog Inputs for Internal Logging of Other Instruments, 0-2.5 V | 3 Inputs | 3 Inputs | No |
Serial Output, RS232 | Yes | Yes | Yes |
USB Output | Optional External Adapter | Optional External Adapter | Yes |
Flash Card Output | Yes | Yes | No |
Data Transfer Baud Rates | 2400, 4800, 19200 | 2400, 4800, 19200 | 2400, 4800, 19200 |
Analog Voltage Output | 0-2.5 V | 0-2.5 V | 0-2.5 V |
Analog Current Output | Optional 4-20 mA | Optional 4-20 mA | 4-20 mA |
Output Ranges | User-Defined Scaling Factor in Menu | User-Defined Scaling Factor in Menu | User-Defined Scaling Factor in Menu |
DewLine? | Yes | Yes | Yes |
Flow Through | No | No | Option |
GPS Option | Yes | Yes | No |
Bluetooth Option | Yes | Yes | Yes |
Relay with 2 Setpoints | No | No | Yes (2) |
Air Pump Lifetime (hours) | 20,000 | 20,000 | 15,000 |
Flow Meter | Yes | Yes | Yes |